Subject: Re: [sloan-25mlog 7541] 2.5m obslog 20241201 (SJD 60646) From: Kaike Date: Mon, 2 Dec 2024 10:27:51 -0700 Hi Jose and Dan, Could it be possible that the recover procedure fails due to a failure = of reading out from camera to netburner? When will the data in the = netburner be cleared (or overridden), in the process of prepare for the = next exposure or prepare for next readout from camera to the netburner? Thanks, -Kaike > On Dec 2, 2024, at 6:22=E2=80=AFAM, Daniel Oravetz via sloan-25mlog = wrote: >=20 >=20 > APACHE POINT OBSERVATORY SDSS 2.5M >=20 > OBSERVING LOG > Sunday December 01, 2024 >=20 > (SJD 60646) >=20 > ---- Observing Team ---- >=20 > Dan Oravetz >=20 >=20 > (3) On 103940-674791-17538, there was a Netburner error, but despite = many attempts at the recovery procedure the exposure could not be = recovered from the daq. >=20 > The science exposure sequence on 103940-674792-17539 aborted in hal = because the previous BOSS exposure's (376995) readout timed-out as a = result of a "sp1daq readout failed: Invalid response from Netburner" = message at 07:22:07Z. However, whenever I tried the usual recovery = procedure (and I tried many attempts at this), the message from BOSS was = that there was no exposure to act on; i.e., nothing in the daq to = re-read or recover. >=20 > When I was checking the status of the controllers, most of the time = all three BOSS controllers replied, but a couple of times sp1daq did not = reply and at least once it actually did show a 'DEAD' status (which = isn't unusual when we have to recover exposures and is actually the norm = when we have a Netburner error). After reloading controllers several = times (which is sometimes necessary during the recovery procedure), and = making sure that there was an 'OK' state for all three BOSS controllers, = there still was "no exposure to act on" when I tried to recover the = exposure. >=20 > Ultimately, I tried all of the 'boss exposure [...]' status commands = -- 'pause', 'readout', 'abort', recover', etc. -- to see if I could = change the BOSS Exp Status and none of them had any effect. When I = eventually tried a 'boss iccStatus' it did automatically force a readout = attempt, which at first seemed hopeful, but was unsuccessful with each = attempt (the readout would just count down to 0.0 and never move beyond = the 'Reading 0.0' state). >=20 > After all of these unsuccessful attempts at recovering the exposure, = all three BOSS controllers were repeatedly replying with an 'OK' state, = but before restarting bossICC as the next step I was resigned to try the = 'boss clearExposure' command (which clears any attempt at recovering the = exposure, even though everything up to this point was unsuccessful = anyway). That command did clear the BOSS Exp Status from the 'Reading = 0.0' state to an 'Idle' state, so I tried a bias and the bias read-out = successfully. I did not see an urgent need to restart bossICC at that = point because the BOSS controllers (and BOSS itself) seemed normally = responsive, so I moved on to the next field. >=20 > I didn't have any problems with BOSS the rest of the night. >=20