Subject: Re: [sloan-25mlog 6559] 2.5m obslog 20230125 (SJD 59970) From: John Date: Thu, 26 Jan 2023 11:54:36 -0700 > On Jan 26, 2023, at 6:22 AM, dmbiz@apo.nmsu.edu wrote: >=20 >=20 > APACHE POINT OBSERVATORY SDSS 2.5M >=20 > OBSERVING LOG > Wednesday January 25, 2023 >=20 > (SJD 59970) >=20 >=20 >=20 > The first set of dome flats shows low throughput in the MTPs 6 and 7: >=20 > MJD DomeFlat MTPs: 1 2 3 4 5 6 7 8 9 = 10 AZ ALT ROT > 59970 apq-44080010.fits 1.01 1.03 1.05 1.05 1.05 0.60 0.42 1.01 0.99 = 0.97 121.0 30.0 0.0 > 59970 apq-44080011.fits 1.03 1.03 1.05 1.04 1.04 0.59 0.42 1.01 0.99 = 0.96 121.0 30.0 0.0 > 59970 apq-44080012.fits 1.04 1.03 1.05 1.04 1.04 0.59 0.41 1.01 0.99 = 0.96 121.0 30.0 0.0 >=20 > I disconnected the gang connector, blowed it, and reseated back, and = took 3 more dome flats > 59970 apq-44080013.fits 1.01 1.01 1.04 1.03 1.03 0.98 0.97 1.00 0.98 = 0.95 60.0 60.0 60.0 > 59970 apq-44080014.fits 1.01 1.01 1.03 1.02 1.03 0.98 0.96 1.00 0.98 = 0.95 60.0 60.0 60.0 > 59970 apq-44080015.fits 1.02 1.01 1.04 1.02 1.03 0.97 0.96 1.00 0.98 = 0.95 60.0 60.0 60.0 >=20 > Now all MTPs look good! It=E2=80=99s unfortunate, but it looks like this is what will be = required on the dome flat side of things. =20 Plug, test, clean, re-test On the good side however, at least it is just once per night instead of = once per cartridge.=