Subject: Re: [sloan-25mlog 5911] 2.5m obslog 20220301 (SJD 59640) From: Ryan Wagner Date: Wed, 2 Mar 2022 08:01:49 -0700 --Apple-Mail=_CC3D2ED6-9C0C-4652-926E-742E50EDD749 Content-Transfer-Encoding: quoted-printable Content-Type: text/plain; charset=us-ascii I have asked James to take look into this.=20 The power source was an issue on the R1 Ion pump a few months ago, when = we had the warmed camera issue, but the problem had vanished.=20 Thank you for pointing this out Dan.=20 Ryan > On Mar 2, 2022, at 05:53, doravetz@apo.nmsu.edu wrote: >=20 > ---- Engineering Issues ---- >=20 > FYI, there was an r1 ion pump warning in the STUI alerts upon arrival = this afternoon. Looking at the Grafana plots, about 06:00Z the r1 ion = pump increased to ~logP=3D-5.4 at 06:10Z, then dropped down below -7 = before rebounding and dropping again. That "burping" event lasted = several hours but then stabilized and has been steady at about logP=3D-6.7= since 18:00Z. --Apple-Mail=_CC3D2ED6-9C0C-4652-926E-742E50EDD749 Content-Transfer-Encoding: quoted-printable Content-Type: text/html; charset=us-ascii I = have asked James to take look into this. The power = source was an issue on the R1 Ion pump a few months ago, when we had the = warmed camera issue, but the problem had vanished. Thank you for pointing this out Dan. RyanOn Mar 2, 2022, at 05:53, doravetz@apo.nmsu.edu= wrote:---- = Engineering Issues ----FYI, there was an r1 ion pump warning in the STUI alerts upon = arrival this afternoon. Looking at the Grafana plots, about 06:00Z the = r1 ion pump increased to ~logP=3D-5.4 at 06:10Z, then dropped down below = -7 before rebounding and dropping again. That "burping" event lasted = several hours but then stabilized and has been steady at about logP=3D-6.7= since 18:00Z.= --Apple-Mail=_CC3D2ED6-9C0C-4652-926E-742E50EDD749--