Date: Mon, 20 Jan 2020 13:03:48 -0000 Subject: Re: [sloan-25mlog 4894] 2.5m obslog 20200119 (SJD 58868) From: "Dmitry Bizyaev" Good morning, the same problem was reported on October 5, 2019. That time it was recommended to notify Ed Leon in addition to 2.5m engineering staff. CC-in Ed. That time Russet commented: >It's not uncommon for the TSL sensors to freeze when the dewpoint is higher >than 0 C. By "freeze" I mean physically, with ice buildup. There is a >procedure for unfreezing it but no one on the night staff is trained in that >procedure. Dmitry > Sunday January 19, 2020 > (SJD 58868) > (1) We are concerned that the humidity sensors may be indicating that it is > safe > to open when it is not. Is there a way to check the accuracy of the dew point > sensors to make sure they are reading properly? See log for details. > NOTE: > ----- > The past few nights the humidity has not been registering very high on the APO > Weather Status page but is clearly seen with frost on the ground, seeing one's > breath, and feeling moisture on surfaces. > > We closed at 5:31Z after noticing high humidity and condensation outside; we > attempted to reopen at 07:15Z (the dewpoint depression was >8C) but > frost/condensation was clearly visible on several surfaces outside (rails near > ops building, walls of the plug lab, rails along the apron, enclosure doors). > There were several times tonight when the dew point depression indicated it > was > very safe to open, but conditions outside indicated that it was not.