Subject: Re: [sloan-25mlog 4762] Re: 2.5m obslog 20191112 (SJD 58800) From: John Downey Date: Wed, 13 Nov 2019 12:55:52 -0700 >=20 >>=20 >>> So looks like we have >>> a full sensor issue with SP1. >> Not this time, it was most likely a leaking quick disconnect fitting. = I would have liked to see some closer inspection to confirm that, = because even if you can't get it connected well enough to fill, it would = be useful information for us when it comes to fixing the connector. >>=20 >> I think the trouble Dylan noted should have been the tip off for = this. When I tried the SP1 fitting, I was able to get it to slip but I = had to turn it really tight to do it, tighter than I usually leave it in = the first place. Given what you're reporting about spills and pressure = problems, I suspect that there is also some connection error going on. >=20 > Again, this should not the LN2 line connection issue. LN2 spilled out = from the 10L valve not from the LN2 line or connection > joint.=20 >=20 As to that, we'll never know unless it repeats. What you described is consistent with the conclusion that it was a = misconnection, especially since the stop did not work. A full sensor = problem would not keep the stop button from stopping the fill.=