From: Karen Kinemuchi Subject: Re: [sloan-25mlog 4208] 2.5m obslog 20180929 (SJD 58391) Date: Sun, 30 Sep 2018 07:03:37 -0600 A quick amendment: > On Sep 30, 2018, at 6:52 AM, viktorm@apo.nmsu.edu wrote: >=20 > NIGHT: > =3D=3D=3D=3D=3D=3D > 00:58:26 encl25m : open; =20 >=20 > 00:24:17 loadCart: Cart 13, Plate 10724; > 00:24:17 survey: eBOSS; lead: None >=20 >=20 > 23:51 Z sdR-b1-00272533.fit.gz: WARNING: Cd I 3610 line missing (lamps = not warm?) > 23:51 Z sdR-b2-00272533.fit.gz: WARNING: Cd I 3610 line missing (lamps = not warm?) >=20 sp2daq failed and only one of the Hartmanns completed. GoToField = command failed at the Hartmann step. BOSS instrument window hung at the = readout of the one Hartmann. To clear it, I issued a =E2=80=9Cboss = ping=E2=80=9D (6 OKs), =E2=80=9Cboss reloadControllers=E2=80=9D for = assurance, =E2=80=9Cboss clearExposure=E2=80=9D to clear the hung state = in the BOSS instrument window, and then a test bias =E2=80=9Cboss = exposure bias=E2=80=9D. =20 Reissued the goToField sequence and commands completed normally. = Hartmanns and cals were completed in two sets before 15 degree twilight. = Sorry no Hartmann outputs here - I=E2=80=99ve logged out of STUI. > 01:29:01 sop.gotoField =3D failed; =20 > 01:31Z Karen did 'boss reload controller' and 'boss clear exposure' = and check > bias. All is ok now. >=20 > 01:54:42 boss exposure 900.0, file=3D272544=20 > 02:11:03 boss exposure 900.0, file=3D272545=20 > 02:27:24 boss exposure 900.0, file=3D272546=20 > 02:43:48 boss exposure 900.0, file=3D272547=20 > 03:00:10 boss exposure 900.0, file=3D272548=20