From: John Downey Subject: Fwd: Particle sampling issue found Date: Tue, 6 Nov 2018 14:00:29 -0700 --Apple-Mail=_2D794EFE-295A-4CF4-91F3-0096D70DF70F Content-Transfer-Encoding: quoted-printable Content-Type: text/plain; charset=us-ascii Big thank you to Ed! > Begin forwarded message: >=20 > From: Ed Leon > Subject: Particle sampling issue found > Date: November 6, 2018 at 1:58:32 PM MST > To: John Downey > Cc: kaike pan >=20 > Gentlemen, >=20 > I did find that the outside sampling tube had been left disconnected = from the last service. This explains the low counts - sorry. =20 >=20 > I did perform the side by side experiment with the hand held particle = counter. Web page displays counts per 0.1cubic ft and the hand held = displays counts per 1.0 cubic ft. Sampling flow rate is the same for = both units (0.1cubic ft per minute). > F.Y.I. >=20 > With the long sampling tube: > hand held web page > 0.3um 128990 4493 > 1.0um 3380 156 >=20 > With the short sampling tube: > 0.3um 127600 9285 > 1.0um 3510 221 >=20 > Samples above were taken within 5 minutes of each other. >=20 > The experiment is VERY uncontrolled so I doubt any correlations can be = made from these two samples. The service counters are on a 5 minute = cadence so a controlled experiment would require software support to = bring the experiment inside. Suffice to say that the particle counters = (not the hand held) are calibrated yearly. The count data can only = provide relative (compared to historical) information at best. >=20 > Ed Leon > Apache Point Observatory > 2001 Apache Point Road > Sunspot, NM 88349 >=20 > 575-437-0552 >=20 --Apple-Mail=_2D794EFE-295A-4CF4-91F3-0096D70DF70F Content-Transfer-Encoding: quoted-printable Content-Type: text/html; charset=us-ascii Big thank you to Ed!Begin = forwarded message:From: = Ed Leon Subject: = Particle sampling = issue foundDate: November 6, 2018 at 1:58:32 PM MSTTo: = John Downey Cc: kaike pan Gentlemen,I did = find that the outside sampling tube had been left disconnected from the = last service. This explains the low counts - sorry. I did perform the side = by side experiment with the hand held particle counter. Web page = displays counts per 0.1cubic ft and the hand held displays counts per = 1.0 cubic ft. Sampling flow rate is the same for both units = (0.1cubic ft per minute).F.Y.I.With the long sampling tube: = hand held = web page0.3um 128990 = 44931.0um 3380 = 156With the short sampling tube:0.3um = 127600 = 92851.0um 3510 = 221Samples above were taken within 5 minutes of each = other.The = experiment is VERY uncontrolled so I doubt any correlations can be made = from these two samples. The service counters are on a 5 minute = cadence so a controlled experiment would require software support to = bring the experiment inside. Suffice to say that the particle = counters (not the hand held) are calibrated yearly. The count data = can only provide relative (compared to historical) information at = best. Ed = LeonApache Point Observatory2001 Apache Point RoadSunspot, NM = 88349575-437-0552 = --Apple-Mail=_2D794EFE-295A-4CF4-91F3-0096D70DF70F--