Subject: Re: [sloan-25mlog 3853] 2.5m obslog 20180125 (SJD 58144) From: John Downey Date: Fri, 26 Jan 2018 10:48:49 -0700 Looking back at the instance of flat field problems on Dec 23/24 2017, = it was also petal #2. The description then was a bit less clear to me, but I would be that it = was the same - one of the switches stuck, confusing the MCP and PLC as = to the state of the petal. These switches are readily available for $14, I believe we will just = replace them all as it does not look too difficult to do. I only have = three spares on hand, so I will probably replace the ones in #2 today, = then do the rest during engineering next week or the following month. > On Jan 26, 2018, at 10:00 AM, John Downey = wrote: >=20 >=20 >> Plate 8201 is not complete. >> 13:10Z We had some trouble moving the telescope to 5 degrees for = closing. The >> goTo5 script was stating that the telescope "did not decelerate" and = stopped at >> alt=3D5.5 degrees. We deemed it unsafe to move the enclosure and = tried moving the >> telescope back to stow and back to 5 degrees. We noticed that the #2 = flat field >> petal was reporting in the MCP window as in a confused state, ie "?". = Checking >> the interlocks for that petal, the interlock stated that Petal #2 was = still >> open. Visual inspection show that the petal was flush against the = telescope and >> closed. >>=20 >> Solution: Open the panel and set to manual mode for the flat field = petals. We >> opened and closed Petal #2. The sensor then read that all the flat = field petals >> were closed. >>=20 >> Re-issued to the command to goTo5 using the script. Success and we = have now closed. >>=20 >> Can someone check the sensor to Petal #2, as well as the motor and = belt? Upon >> visual inspection, the belt looked ok. With auto-close of the = petals, the MCP >> was reporting a confused and/or state for this petal, even though it = looked like >> it was closed. =20 >=20 > Good job figuring it out, cycling the petal manually is the first = thing to try. >=20 > The way the system works, the open and closed limit switches also act = as the open and closed sensors. Given the problem and the means by = which it was corrected, I suspect that the mechanical switch for "open" = had stuck and the act of manually moving the petal from closed to open = caused it to unstick. >=20 > Probably we need to look at replacing all the switches - they have = been in use about 4 years now with some exposure to dust and moisture, = probably we're starting to see the limit of their maintenance free = lifespans. >=20 > If its easy to do I'll order switches and take care of it during an = engineering week, if it requires a great deal of disassembly it will go = on next summer's list. >=20 >=20 > ____________________________________________________________ > Logs of Nightly APO Activities > = https://trac.sdss.org/tracmailman/browser/private/sloan-25mlog/index.html