Subject: Re: [sloan-25mlog 3599] 2.5m obslog 20170705 (SJD 57940) From: John Downey Date: Thu, 6 Jul 2017 05:38:54 -0600 >=20 >=20 > ---- Observing Summary ---- >=20 > Lost 36 minutes to hardware. > The primary and secondary fills did not start as expected, since the = toggle > switch for SP2 on the silver autofill box was turned off with the R2 = warming > today. Further details in log. >=20 This was my fault - used the toggles to repressurize the 10 liter when = working on warming R2 (fill line to R2 disconnected and capped, lost = pressure while doing this). The locking toggles seem a bit floppy to me, we may need to replace = them.