Subject: Re: [sloan-25mlog 3115] Re: 2.5m obslog 20160923 (SJD 57655) From: Dan Long Date: Sat, 24 Sep 2016 11:39:08 -0600 > On Sep 24, 2016, at 10:35 AM, jdowney@apo.nmsu.edu wrote: >=20 >=20 >>>=20 >>> (3) At 07:59Z we had to interrupt the night to force a secondary = fill >>> because >>> the SP2 10L dewar had run dry. The fill took 30 minutes to complete. = See >>> Log for >>> details. >>=20 >>=20 >>> 02:07Z - Opened. Recommenced the force-fill of the BOSS = spectrographs. >>>=20 >>> 02:45Z - After almost 40 minutes since restarting the secondary = fill, >>> the SP2 >>> secondary fill was still in progress. The TPM was reporting the SP2 = 180L >>> LN2 >>> pressure at 10 psi (rather low). A total of ~25 lbs of LN2 = transferred >>> from the >>> SP2 180L dewar during the two separate evening fills (the fill we >>> started while >>> we were on the phone with JohnD and DanL was the first fill which we >>> eventually >>> stopped to troubleshoot the enclosure, and the fill we started after = we >>> finally >>> got the enclosure off was the second fill). Last night, the SP2 >>> secondary fill >>> took 50 minutes to complete and 15-20 lbs of LN2 transferred, so = since >>> ~25 lbs >>> of LN2 had transferred tonight and the fill was still going while we = had >>> now >>> entered dark time, the decision was made to stop the ongoing = secondary >>> fill and >>> start science since the sky was clear. >>=20 >> Looks like that the low pressure on 180L dewar was the root of the >> problem. If there >> is no way to rise the pressure when a secondary fill is needed, we = should >> consider to >> swap out the reward. At the mean, we will discuss with day crew on = the >> pressure >> issue on these dewars. >>=20 >=20 > Tracy said that it probably didn't have time to build to full pressure > after he replaced the popoff valve. Next time this happens we'll just > replace the dewar instead, or use another to bump the pressure up. >=20 > John >=20 If the dewar gets below 15 PSI, fills will become inefficient and time = consuming. At around 10 PSI, one starts seeing problems like low dewar = pressure warnings and primary fill faults (should a primary fill occur = when the 10L is still at very low pressure). The bottom line is a 180L = dewar should be changed for another if it can=E2=80=99t maintain = pressure above 15 PSI. =20 -Dan > ____________________________________________________________ > Logs of Nightly APO Activities > = https://trac.sdss.org/tracmailman/browser/private/sloan-25mlog/index.html