Subject: Re: [sloan-25mlog 2863] 2.5m obslog 20160304 (SJD 57452) From: Karen Kinemuchi Date: Sat, 5 Mar 2016 15:50:20 -0700 Hi Dan, Yes, I remember this quick J/K test to check the counterweight = balance. I can run that this afternoon during checkout (it's cloudy = now with radar in the county) and can report the values I find in the = night log. If we find the numbers to be grossly out, we can reset the = counterweight values to the ones posted on the White Board of Powerful = Info in the control room. Thanks! Karen and Audrey On Mar 5, 2016, at 8:15 AM, Dan Long wrote: > Hi Karen and Audrey, >=20 > That is definitely an indication that the altitude is out of balance = and top heavy. There is a test that once upon a time was part of = afternoon checkout, where one used the mcp j and k incremental velocity = keys to look at the motor voltages going up and down slowly (5000 = cts/sec or so). =20 >=20 > One should see a fairly symmetric reading of about 1000 units going = up and -1000 units going down when monitoring over several seconds as = seen in the mcpMenu. My guess is the counterweights are around 20-30 = ticks too high (the dewars between full and empty will change the = balance on the order of 10-15 ticks). >=20 > Cheers, >=20 > Dan >=20 >=20 >> On Mar 5, 2016, at 6:01 AM, kinemuchi@apo.nmsu.edu wrote: >>=20 >>=20 >> APACHE POINT OBSERVATORY SDSS 2.5M >>=20 >> OBSERVING LOG >> Friday March 04, 2016 >>=20 >> (SJD 57452) >>=20 >> ---- Observing Team ---- >>=20 >> Audrey Oravetz >> Karen Kinemuchi >>=20 >> kaike Pan (afternoon) >>=20 >>=20 > . > . > . >=20 >> When the telescope is slow to the stow position, 121, 30. It stopped = at >> 121, 30. But as soon as "axis stop" and a stop was engaged, it went = dow >> 0.5 degree to 29.5. When LN2 was connected, it started filling. It = was a little >> top-heavy.=20 >>=20 >> APO evening calibration was taken, checked. >>=20 >> 00:15Z started ventilation. >>=20 >> NIGHT: >> =3D=3D=3D=3D=3D=3D >> 01:08:55 encl25m : open; =20 >>=20 >> We noticed that after the telescope did the primary fill and moving = from 5 >> degree to stow, after the issuing of tcc axis stop and stop button, = the >> telescope still slipped by 0.5 degree in altitude (30 degrees -> = 29.5). We >> think it may not be related to the dewars being empty or not. We = will continue >> to monitor the telescope and see if more altitude slips happen around = stow position. >>=20 >=20 > . > . > . >> ---- Telescope Status ---- >>=20 >> 00:15Z Started ventilation >> 01:08Z encl25m : open >> 12:44Z encl25m : closed >>=20 >> No interlocks bypassed >>=20 >> Status at: 12:55:38=20 >> Telescope stowed at: 121.0, 29.6, 0.0 mount >> Instrument mounted: cartridge =3D 3 >> Counterweights at: 454, 455, 455, 455=20 >> LN2 autofill systems: Connected and turned on >> 180L LN2 dewar scales: SP1 209.8 lbs, 16.2 psi >> SP2 232.0 lbs, 20.2 psi >=20