Subject: Re: [sdss3-25mlog 1330] Re: 2.5m obslog 20130718 (MJD 56492) From: Kaike Pan Date: Fri, 19 Jul 2013 16:50:17 -0600 On Jul 19, 2013, at 9:19 AM, Dan Long wrote: > Hi Dmitry, >=20 > Just to let you know, the secondary dewar (10L) pressure being zero = will not be fixed by startfillseq. That command requests the camera = dewars to fill from the 10L dewars. If the pressure is zero in the 10L = the primary fill will fail. Hence the second alert of SP1LN2Fill Fault. = =20 >=20 > Forcing a fill on the 10L dewars by pushing the fill button on the = silver autofill control box is the way to fix the problem if the reason = for zero pressure is an empty 10L. =20 >=20 > With the lines hooked up it shouldn't have gone empty unless the = sensor didn't trigger or the ground connection was bad. Since you were = able to force a fill on the 10L, it probably wasn't the ground. It is = strange that it would have gone empty since the interval between = secondary fills wasn't abnormally long (about 13 hours), so maybe the = dewar pressure going to zero wasn't because it was empty (but it is the = usual cause). So the reason for the zero pressure on sp1's 10L is a bit = of a mystery. >=20 >=20 > Cheers, >=20 > Dan >=20 Sometimes, an extremely low pressure is seen after a primary fill is = just finished and there is no more N2 in the 10L dewar. -kaike =20=