Subject: Re: 2.5m obslog 20100608 (MJD 55356) From: Wendell Jordan Date: Wed, 9 Jun 2010 09:40:02 -0600 Connie, I did a couple of startfillseq commands to sp1 and sp2 alternately. Sp1 = does NOT do the added "timed insurance fill" following its initial fill, = however sp2 DOES give the added "timed insurance fill" following its = initial fill. Liquid LN2 does spew out during sp2's insurance fill, but = liquid LN2 does not spew out of sp1 cameras that I could see (just the = gas). Again, sp1 does not do an insurance fill--Could this lack of = timed insurance topping off be part of sp1's problem? Both sp1 camera valves seem to behave correctly during commanded fills, = and turn off within several seconds of one another. I still plan to do = a coil resistance check however. Wendell On Jun 9, 2010, at 8:35 AM, Constance Rockosi wrote: >=20 > Looks like sp2 red got its ln2 dumped out. =20 >=20 > David -- you are correct about the dictionary error. But > to really answer you question I need to know where the data > comes from. ln2check, tcheck, or showreadvolts? >=20 > --connie >=20 > On Wed, Jun 09, 2010 at 07:31:07AM -0700, David Kirkby wrote: >> The SP2R0 LN2 temperature is spiking right now but was fine earlier = (see the >> attached plot of LN2 temps over the past 3 days). Is that expected? >>=20 >> The BOSS dictionary claims that it reports LN2 temperatures in Volts = but >> that doesn't sound right looking at the graph. >>=20 >> David >>=20 >> On Wed, Jun 9, 2010 at 4:40 AM, Constance Rockosi = wrote: >>=20 >>>=20 >>> There is some padding that lets the CCDs warm up a little bit before >>> you get a temperature error, so I am not very worried that you = didn't >>> get a temperature error. >>>=20 >>> You should have gotten a fillfault error. Did you? >>>=20 >>> Ah. If you didn't, then that might be telling us that the problem >>> is that the full sensor has failed open, so that the camera micro >>> always sees the dewar as full. >>>=20 >>> --Connie >>>=20 >>>=20 >>=20 >>=20 >> !DSPAM:10028,4c0fa5ad5231021468! >=20 >=20 >=20 > --=20 >=20 > Associate Professor, Assoc. Astronomer > UC Santa Cruz, UCO/Lick Observatory > Office: 831 459 5246 > crockosi@ucolick.org > http://www.ucolick.org/~crockosi >=20 > DSS DSS DSS DSS DSS Digital Sky Survey DSS DSS DSS DSS DSS DSS > DSS *** The information posted to SDSS mailing lists is proprietary = *** > DSS This is message 5250 in the sdss-25mlog archive, URL > DSS = http://www.astro.princeton.edu:81/sdss-25mlog/msg.5250.html > DSS The index is at = http://www.astro.princeton.edu:81/sdss-25mlog/INDEX.html > DSS To join/leave the list, go to > DSS = http://www.astro.princeton.edu:81/cgi-bin/mailinglists.pl/show_subscriptio= n?list=3Dsdss-25mlog > DSS To post a message, mail it to sdss-25mlog@astro.princeton.edu > DSS > DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS DSS