From: stephanie snedden Subject: Re: 2.5m obslog 20080116 (MJD 54482) Date: Thu, 17 Jan 2008 11:30:03 -0700 --Apple-Mail-5--1050895324 Content-Transfer-Encoding: 7bit Content-Type: text/plain; charset=US-ASCII; delsp=yes; format=flowed This is the same problem that we had recently (in the last month, anyway), where if the spectro focus is set for a cold temperature, the collimator actuators reach a limit during the dithered flat tilt moves and fail to complete the move. This was a topic in the last ops phone con, and the solution is to set the spectrograph red camera lenses to a "winter position" so the typical winter focus is not so near the actuator limit. We have no written procedure for this, although I hope to have a draft ready for testing this shakeup/down. For tonight and tomorrow night, if the focus is set for -2C or colder and the temps are really that cold, don't bother trying dithered flats. Steph On Jan 17, 2008, at 4:44 AM, Audrey Simmons wrote: > > > We attmepted to gather dithered flats several different times, but > an error > repeatedly came back saying that the mechanicals did not respond > correctly. > Maybe the motors had trouble because of the cold temperatures? > > 08:50Z - Took dithered flats, starting with exposure 49049 and > ending at > 49058. After the tenth flat, and error message came back: > > "Finished with specFlat 10 of 15. > Setting up Collimator > 2008-01-17 09:30:41Z: Original command was !mb100! > 2008-01-17 09:30:41Z: expected reply was !mb100 > OK! > motor b: Mechanicals did not respond correctly; response was mb100 > > invalid keyed list format: list contains an empty field entry > Error: problem with motors (see previous error message) > Original motor positions: {a -15424} {b -16010} {c -16610} > Current motor positions: {a -15313} {b -15898} {c -16610}" > --Apple-Mail-5--1050895324 Content-Transfer-Encoding: quoted-printable Content-Type: text/html; charset=ISO-8859-1 This is the same problem that we had recently (in the last month, = anyway), where if the spectro focus is set for a cold temperature, the = collimator actuators reach a limit during the dithered flat tilt moves = and fail to complete the move. This was a topic in the last ops phone = con, and the solution is to set the spectrograph red camera lenses to a = "winter position" so the typical winter focus is not so near the = actuator limit. We have no written procedure for this, although I hope = to have a draft ready for testing this shakeup/down. For tonight and = tomorrow night, if the focus is set for -2C or colder and the temps are = really that cold, don't bother trying dithered flats.StephOn Jan 17, 2008, at = 4:44 AM, Audrey Simmons wrote:We = attmepted to gather dithered flats several different times, but an = error=A0repeatedly came back saying that the mechanicals did = not respond correctly.=A0Maybe = the motors had trouble because of the cold temperatures?08:50Z - Took dithered flats, = starting with exposure 49049 and ending at=A049058. = After the tenth flat, and error message came back:"Finished with specFlat 10 of 15.Setting up = Collimator2008-01-17 09:30:41Z: Original = command was !mb100!2008-01-17 09:30:41Z: = expected reply was !mb100OK!motor b: Mechanicals did not respond correctly; = response was mb100invalid keyed list format: list = contains an empty field entryError: = problem with motors (see previous error message)Original motor positions: {a -15424} {b -16010} {c = -16610}Current motor positions: {a = -15313} {b -15898} {c -16610}"= --Apple-Mail-5--1050895324--