From: Jon Brinkmann Date: Thu, 28 Apr 2005 14:20:16 -0600 Subject: Re: specMech reverted to previous version: 2.5m obslog 20050427 (MJD 53488) I've run tests on the reverted specMech code. I first tested by telneting to each spectrograph. I did simple exposures, paused and resumed exposures, and aborted exposures. I also tested the colimator motors, Hartmann screens, and shutter. All worked well. I then started SOP and the SOPGUI. I did three precals (flats and arcs) followed by an assortment of science exposures: simple, paused and resumed, and "endStare"d early. I also did three science exposures and three darks using the "goStare command" from the command line. All have correct exposure times in the FITS header. Jon