From: Jon Brinkmann Date: Wed, 21 Apr 2004 11:47:25 -0600 Subject: Re: Spectro focus once again [was RE: 2.5m obslog 20040419 (MJD 53115)] Where are the collimator numbers for the late-night focus? Also, what was adjusted? Was it only the piston on sp1, as per David's re-run of the idlspec2d collimate procedure? > From http://sos.apo.nmsu.edu/logfile-53115.html, Tair = 12.9C for the > preflight focus check exposures, but by the time Wsigma went out of tolerance, > Tair = 4.3C, so deltaT = -8.8C. This is non-trivial but not, I think, > particularly large for this site. But it is the spectrograph body temperature that matters, not the air. The spectrograph bodies are not ventilated or actively cooled. We use the PMSS average temperature as an estimate of the body temperature. Note that sp1 needed to be pistoned in all three actuators. That doesn't look like an actuator problem. Also, there were no reported problems moving the mirror to obtain focus, which would indicate an actuator problem. An actuator coupling problem can't cause a focus shift. It will only cause problems moving the mirror. I suspect that the spectrograph body was warmer than the air at opening and slowly cooled during the night, eventually requiring recollimation. I inspected all the bolts that I can check without disassembling actuators. Everything is tight. I didn't touch the mirror positions. Also, you are supposed to be recording temperatures when you do focus checks as so: im> caget tpm_TM_PMSSAvg tpm_TM_TelairAvg.VALA tpm_TM_PMSSAvg 5.48071 tpm_TM_TelairAvg.VALA 6.75 The last time temperatures were recorded in the log was March 24th. You are also to check the slithead latch pressure on the gauge on sp2 near the slithead door. I checked it today. It was 105 psi. Jon