Date: Fri, 20 Feb 2004 09:47:22 -0700 From: Dan Long Subject: Re: 2.5m obslog 20040219 (MJD 53055) > > OBSERVING LOG >> Thursday February 19, 2004 >> >> (MJD 53055) > ... >> Plate 1461 (cartridge 2): FK5 again needed to acquire the field. The FK5 >> was 10" S of its original pointing. After the preCal finished it >>was noticed >> that sp1 had gone out of collimation (r1 XSIGMA and WSIGMA were both in the >> red). The science exposure was aborted at this time and sp1 was >>recollimated. >> The time lost to recollimation was logged as a spectrograph problem. > >Shouldn't we have a separate logging category for spectrograph focussing? > >Jon Hi Jon, I think this is appropriate. The spectrographs were focused in the afternoon and the night time temperature dropped less than 5 C. Thermal issues shouldn't play a part with such small temperature change. Since refocusing was required it would indicate that there is a problem, especially since it only went out of spec in sp1. Cheers, Dan