From: Jon Brinkmann Date: Sun, 3 Mar 2002 09:11:55 -0700 Subject: Re: 2.5m obslog 20020302 (MJD 52336) What a night!!!!! FYI, the specmech code doesn't start timing an exposure until it sees that the Hartmann doors and shutter are open. That is why it wouldn't start counting: it was waiting for the shutter to open. I exercised both spectrographs this morning. Spectrograph one appears to be operating properly, but spectrograph two's shutter will open but not close completely. Both Hartmann doors appear to be working properly. Spectrograph two may still have a small air leak, something may have gotten caught in the shutter or the closed sensor or wiring may have failed. I will be on site around 2 PM to physically check and fix it. Jon