From: "Pete Newman" Subject: Watcher messages [was RE: 2.5m obslog 20011110 (MJD 52224)] Date: Sun, 11 Nov 2001 15:17:51 -0700 > Some other messages in Watcher that are new and we would like to know > how to act upon seeing them: > Inconsistent instrument ID switches: 6 14 14 (Inconsistent instrument ID > switches: 6 14 (Nov 11 02:34:09): Nov 11 02:34:19 MST) This one seems to happen frequently as a cartridge is mounted, but before all the switches are properly engaged. Ignore it. > Text="pointing error probe does not exist:check inst data": (Nov 11 > 02:51:57 MST) This one comes from TCC whenever a track /pterr command is issued (I believe, as in fk5InFiber), because there is no guider instrument block for the spectrograph instrument (that's what the pointing error probe bit means). Ignore it. > Modu="prt GetChan" (Nov 11 02:45:29 MST) > Text="could not obtain guide camera I/O channel":: (Nov 11 02:45:29 MST) This one is new to me, too. --Pete